KOH ETCHING
PROCEDURE
This process is to be done at
the wbgeneral. There is a temperature controlled water bath set up.
1) Fill it with water (a couple of
inches from the top so that it does not overflow when the beaker is in
it).
2) Turn it on with the on off
toggle and set the temp to 80C. Set the Limit Adjust knob at about 130C (if it
is too low it will trip the breaker and cool down the bath. After about 3 hours
check the water level and add water to the bath if needed.
3) The KOH beaker with your desired
percentage of KOH should be immersed into the temperature controlled bath. Place
your wafers in either the designated Teflon boat or single wafer Teflon holder
and put them in the solution.
4) Place the condensing unit on top
of the beaker with the lower hose connected to the industrial water connection
and the upper hose going down the drain. Turn on the water (slow steady flow). A
thermometer can be placed in the top to monitor the temperature of the bath.
NOTE: Most people use 30% KOH
(dilute the 45% solution that we buy to be 30%). It etches at ~60 microns per
hour.
I. LONG ETCHES
- DO NOT LEAVE THE BATH for more
than 4 hours (NOT OVERNIGHT!!) It will boil dry.
II. PATTERNING
- Resist should NOT be put in the
KOH (it will just lift off). You need to use nitride (as little as 250A will
hold up for etching completely through a wafer). Oxide can be used for short
etches, but is not recommended because it etches fast in KOH (rate unknown).
If you want to use oxide as a mask, TMAH is recommended instead of
KOH.
III. STORAGE and
REUSE
- You may save your solution in in
a plastic bottle (cool it first) as long as it is properly labeled and in a
secondary containment (i.e. the bottom of the cart).
IV. DESIGNATED KOH BEAKERS and
HOLDERS
If you plan to do processing in
mainline equipment (i.e.: tylan 1-4, tylanpoly, tylannitride, tylanbpsg
(LTO), drytek2, amtetcher, P5000, lampoly, etc) you MUST CLEAN your
wafers first to remove all KOH contamination.
You MUST have used the
designated quartz KOH etch beaker and the designated "clean" Pyrex
rinse beaker. If you have used any other labware, you may NOT
follow this procedure and your wafers may NOT go in any mainline
equipment. If your wafers have ever had metal on them , they may
not go into the designated quartz ware or into mainline
equipment.
KOH contamination ruins devices
(it is a lifetime killer). Other lab users may have spent months working on
their device wafers. Their work may be ruined if you do not carefully follow
these procedures.
- USE DESIGNATED KOH quartzware
and Teflon holder ONLY.
- If they need to be cleaned, use
room temperature solution of 1:1:1
- H2O:H2O2:HCl
for 20 minutes and then rinse well.
- The KOH quartzware is not to be
used with metals or nonstandard materials because wafers that get KOH etched
often go back into the main stream processing equipment. BEFORE THEY DO they
MUST get a SPECIAL KOH REMOVAL step(see below)
- DO NOT USE GENERAL WETBENCH
beakers and holders etc... because if they are not designated for KOH they
could have been used for anything -- including gold.
V. KOH CLEANUP (removal of KOH
contamination)
If you have used designated KOH
beakers and want to do standard processing in mainline
equipment (i.e. tylan 1-4 and poly and nitride and lto, etc..) you
MUST clean your wafers first to remove KOH contamination.
This may be done in your own
nonmetal contaminated quartzware (specifically used for this clean step
OR with permission you may use the silicide wetbench) (see
Process Staff before doing this).
Procedure at
wbsilicide:
Do not start this procedure
unless you have enough time to completely finish it in one
session, read the whole procedure before you start.
Put a note on the splash shield
that states:
- "KOH Contamination clean-up in
progress in the HCl hot pot"
- your name
- the date
- the time
- the time you expect to be
finished using the wetbench
Be sure to log the clean-up on the
chemical change sheet. Note the fact that the silicide cassette and handle were
both cleaned, too.
CLEAN THE WAFERS:
- Aspirate the current HCL
solution. If the current solution is hot, you must wait until it has
cooled to 50C before aspirating.
- Pour fresh 5:1:1
(H20:H2O2:HCl) solution
- Heat the fresh solution to 70C
- Use your KOH contaminated
tweezers, to put the wafers in a silicide cassette. Put the KOH tweezers in
the KOH tweezer box. Put the box away. You will not use the KOH tweezers after
this point in the clean. Put the KOH contaminated cassette away also.
- Do NOT put the silicide
cassette , with the KOH contaminated wafers, in any dump rinser or hot pot
before going in the 5:1:1 (H2O:H2O2:HCl) bath! (DO NOT DI RINSE
FIRST -- that would contaminate the dump rinser!!)
- Immerse wafers in a
silicide cassette into the 5:1:1
(H20:H2O2:HCl) for at least 20 minutes
heated to 70C.
- After the 5:1:1 step, Dump
rinse the wafers and spin them dry.
- Use non-metal photoresist
tweezers to remove the wafers from the silicide cassette.
- Store the wafers in a non-metal,
non-KOH, contaminated storage box.
- The wafers are now clean enough
to go to wbnonmetal, then wbdiff.
CLEAN THE HOT POT, CASSETTE, and
HANDLE (you must do this immediately after cleaning your
wafers...you may not come back later and clean the wbsilicide
equipment!):
- Aspirate the KOH contaminated
5:1:1 (H2O:H2O2:HCl). Remember, the solution must be cooled to 50C
before aspirating.
- Fill the hot pot (nearly to the
top, so the whole pot is cleaned) with 1:1:1 (H2O:H2O2:HCl).
- DO NOT HEAT the solution (it
will heat up on its own).
- Put the cassette and handle you
used in the solution so that it will will be cleaned at the same time. Let the
solution sit for at least 30 minutes.
- After the 30 minutes, allow the
solution to cool to 50C, then aspirate.
- Rinse the hot pot
thoroughly:
- fill the hot pot to the top with DI water
- aspirate
- fill the hot pot to the top with DI water
- aspirate
- fill the hot pot to the top with DI water, let it overflow for 2 minutes
- aspirate
- fill the hot pot to the top with DI water, lit it overflow for 2 minutes
- aspirate
- Dump rinse and spin dry the
cassette
- Put the cassette handle in with
the other handles
- Pour fresh 5:1:1 (H2O:H2O2:HCl)
solution in the hot pot..
VI. KOH NOTES
- 30% KOH etches at
~1u/min
- Openings >1mm will allow the
KOH V etch to go all the way through a wafer of standard thickness.